Purpose: | Moisture Permeability Of Mask | Test Height: | 150±2mm |
---|---|---|---|
Funnel Height: | 195±10mm | Funnel Diameter: | 150±5mm |
Time Precision: | 0.1s | Control Method: | Manual |
High Light: | 0.1s Moisture Permeability Tester,150mm Moisture Permeability Tester,0.1s Medical Test Equipment |
Time Precision 0.1s Moisture Permeability Tester Of Surgical Masks
Moisture permeability tester
Test standard:
Surgical maskz
Synthetic blood penetration test for surgical masks synthetic blood penetration test for surgical masks can output test reports and store test data. GB 19083-2003-2011 GB/T 4745 requirements.
Characteristics of the equipment
1. Configure funnel for detection.
2. The height can be adjusted within a certain range.
Technical parameters of detector
1. Test height: 150±2mm.
2. Funnel height: 195±10mm.
3. Funnel diameter: 150±5mm
Technical parameters
Product name | Surface moisture tester | |
Fundamental | add water through the funnel and then carry out leakage detection of the measured part | |
Diameter of the funne | 150±5mm | |
Height of the funnel | 195±10mm | |
Test height | 150±2mm | |
Test medium | Water | |
Control method | Manual | |
Scope of application | Surgical mask |
Protective clothing synthetic blood penetration tester
Test standard:
Synthetic blood penetration tester for protective clothing synthetic blood penetration tester for surgical masks can output test reports and store test data. Meet the test standard GB 19082-2010 surgical mask.
Technical parameters
Product name | Synthetic blood penetration tester | |
Fundamental | To simulate breathing through the vacuum pump, and to detect the pressure difference | |
Test pressure | 0-20KPa | |
Micromanometer range | 0-0.1KPa | |
Pressure value resolution | 0.1kPa | |
Time precision | 0.1s | |
Test medium | Synthetic blood | |
Control mode | manual | |
Amount of blood | 50-55ml | |
Scope of application | Protective clothing |
Configuration list
Name | Unit | Quantity | Remarks |
Pressure sensor | unit | 1 | 佛山一众 |
Case | pcs | 1 | Spraying case |
Posstive pressure pumbs | pcs | 1 | Provide compressed air |
Neddle valve | pcs | 2 | Adjust the flow |
Electric appliance | set | 1 | Chin |
Air pumb | unit | 1 | Chint Group |
Pressure relief valve | unit | 1 | Manual pressure relief |
Tank stores | pcs | 1 |
Store compressed air
|
The flame retardant tester for surgical masks is mainly used to test the burning performance of surgical masks after they touch the flame at a certain linear speed.
Characteristics of flame retardant detector for surgical masks
1. The mask fixture is a metal head mold, which can fully simulate the actual use of the mask.
2. The burner is equipped with propane (liquefied gas) and other combustible gas interfaces.
3. The burner height is adjustable.
4. Automatic timing and positioning of burner.
5. Combustion time and flame retardant time are recorded automatically and displayed numerically.
6. Equipped with flame temperature measuring probe and flame temperature display.
Technical specifications of flame retardant detector for surgical masks
1. The linear velocity of the nose * is (60±5) mm/s
2. Flame height can be adjusted (set at 40±4mm before delivery)
3. The diameter of the flame temperature measuring probe is 1.5mm
4. The temperature of the flame at the height (20±2) mm from the top of the burner is (800±50) ℃
5. The distance between the top of the burner and the lowest point of the mask is (20±2) mm
6. The height adjustment range of burner is 30mm
7, metal head die wall thickness 2 ~ 3mm, flame resistant combustion resistant
8. Timer for continuation time and smoldering time: 0 ~ 99.99 seconds, accuracy: ±0.01 seconds
9. Power supply: AC220V, 50Hz, 50W
10. Weight: about 30kg
Applicable standards
Technical requirements for surgical protective masks
YY 0469-2011 surgical masks